We offer high efficacy Oxygen Supply System, which is highly acknowledged in the national as well as international markets. We have firmly positioned ourselves as one of the overruling central pipeline for Oxygen Supply System in India. Made as per international standards, our Oxygen Supply System is highly demanded due to its distinctive features mentioned below :

  • Twin Banks Cylinder Oxygen Manifold suitable to withstand a pressure of 150 Kg/cm2 High pressure copper annealed tail pipes with end Brass adapter suitable for Oxygen Cylinders and manifold
  • Top frame comprising of high pressure copper pipes of size 1/2â�� I.D. x 15swg high pressure brass fittings made of high tensile brass and connections through non- return valves ; high pressure copper tail pipes, made of high pressure copper pipe of size 1/4â�� I.D. x 15 swg
  • The middle and bottom frame fits both round and flat bottom cylinders safely
  • The manifold is tested (hydraulically) at 3500 psig
  • Fully / Semi Automatic Control Panel with two first stage regulators, each capable of delivering 120 psig outlet pressure
  • The First stage regulator connected to the second stage regulator, delivers an outlet pressure of 60 psi. The First two regulators meant for first stage are capable of switchover system incorporated from â��Runningâ�� to â��Reserveâ�� bank due to differential pressure
  • The Control Panel provided for two individual content contact pressure gauges to indicate the cylinder pressure in the two wings of the manifold and common pressure gauges to indicate the delivery/line pressure
  • Control panel has audio alarm for bank changeover
  • The control panel has built in signal lamp indications for changeover and covered with suitable cover indicating the respective services.